Accessibility checklist

One page, run against one screen, before it ships. Accessibility argues the position; this is the list you actually work through, with the item that fails pointing at the page that explains why.

Stable · v1.0 WCAG 2.2 AA EN 301 549 IEC 62366-1
What this is and is not

This is a design and build checklist, not a conformance audit and not a substitute for testing with disabled users. Passing every line here is consistent with a screen that is still unusable, which is why the last section is the only one that measures anything. An accessibility conformance report is a separate exercise with a separate method.

How to run it

Against one screen in one state, not an application. Most of these fail per-state rather than per-screen: the empty variant, the error variant and the alarm-active variant of the same page fail differently, and a checklist run against the happy path finds almost nothing.

Four passes, in this order, because each one is cheapest before the next: keyboard only, then 400 % zoom at 320 px, then greyscale, then a screen reader.

1 · Keyboard only — unplug the mouse

CheckSpecified in
Every control is reachable and operable by keyboard, including anything revealed by hover — SC 2.1.1 Interaction states
Focus is always visible, at 2 px with 2 px offset, and never removed for aesthetics — SC 2.4.7, 2.4.11 Interaction states
Focus order follows reading order; a dialog does not focus its affirmative control Dialog
No keyboard trap; Escape leaves every overlay — SC 2.1.2 Menu · Dialog
Nothing is lost by tabbing away mid-entry; paste is never blocked in any field Sign in
Every gesture has a visible single-pointer equivalent — SC 2.5.1 Touch & gestures
Dragging is never the only means of setting a value — SC 2.5.7 Touch & gestures
Actions fire on release over the target, not on touch-down — SC 2.5.2 Touch & gestures

2 · 400 % zoom, 320 px wide

CheckSpecified in
No loss of content or function; no two-dimensional scrolling except for genuinely two-dimensional content — SC 1.4.10 Grid & layout
Wide content — tables, charts, timelines — scrolls inside its own container; the page body never scrolls horizontally Grid & layout
Text spacing can be overridden without clipping — SC 1.4.12 Typography
Nothing that must not be missed is clipped, truncated or wrapped away — check the patient identity strip specifically Patient header
Targets are at least 44 px, 56 px where gloved or destructive, with dead space between — SC 2.5.8 Touch & gestures
Text is at or above the floor size; nothing carrying a decision is below the decision size Scaling & density
Sticky headers and alarm regions do not eat the viewport at 400 % Elevation

3 · Greyscale and contrast

CheckSpecified in
Every state is still distinguishable with colour removed — alarm priority, selection, validity, changed-vs-unchanged — SC 1.4.1 Colour
Text meets 4.5:1, large text 3:1 — SC 1.4.3 Tokens
Control boundaries and state indicators meet 3:1 — SC 1.4.11. Measure them; this is the check most often assumed rather than run, and this system has failed it twice on its own pages Colour
Alarm colours are unchanged between light and dark themes Tokens — enforced by check-tokens.mjs
Run it in dark mode as a separate pass. Every colour defect this system has shipped was theme-specific Colour
No information carried by position alone in a chart; series are directly labelled Data visualisation

4 · Screen reader

CheckSpecified in
Landmarks, one h1, headings in order, and the page is navigable by heading Accessibility
Tables are real tables with scoped headers; lists are real lists, so announced counts are right Data grid · List
Every control has an accessible name that says what it does; no set of identical link names Button
Alarm conditions are assertive live regions; everything else is polite Messaging
Errors are associated with their field and state a remedy — SC 3.3.1, 3.3.3 Text field
An absent value is announced as absent, distinctly from zero Key–value pair
Charts and image fields carry a description stating quantity, direction and magnitude — a name alone is not a description Data visualisation
Machine-composed text announces its status before the text, not after Authorship
Nothing is announced only in a tooltip; tooltips describe, never label Tooltip

5 · The clinical additions

Not in WCAG. They come from the same place the rest of this system does — a clinician, in a hurry, on a device that was not chosen for them.

CheckSpecified in
No time limit anywhere a user can fail one — SC 2.2.1, and treated as a hard rule rather than a with-exceptions one Lockout & recovery
Motion is suppressed under reduced-motion, not merely slowed; nothing animates routinely Motion
Nothing that must not be missed sits behind a collapse, a hover, a filter or a non-first tab Messaging
Every count has its denominator; nothing is capped Badge
Units are on the surface, never hover-only, never inferred Key–value pair
The screen is legible at the longest sightline it will actually be read from Scaling & density
It works one-handed, in gloves, on a moving cart Touch & gestures

6 · The only part that measures anything

Everything above is inspection, and inspection finds the defects you already know how to look for. The list has a ceiling and this is where it is.

Where this sits in your file

Accessibility failures are use errors. They belong in the usability engineering file (IEC 62366-1) and, where they can lead to harm, in the risk file (ISO 14971) — not in a separate accessibility document nobody reads alongside them. In the EU, EN 301 549 is the harmonised standard that points at WCAG; the accessibility requirements of the European Accessibility Act may also apply depending on your product and users. Paraphrase, not legal advice.

Do's and don'ts

Do

Run it against the error state and the alarm-active state of the screen, separately.

These fail differently from the happy path, and they are the states a clinician is in when the stakes are highest.

Don't

Run it once, on the populated, everything-working view.

The version of the screen nobody has trouble with. It will pass, and it will have told you nothing.

Do

Measure the 3:1 on control boundaries.

This system asserted it twice and was wrong twice — 1.71:1 on a pill border, 1.29:1 on a citation control. Both looked fine to the person who drew them.

Don't

Tick 1.4.11 because the border is visible to you on your monitor.

Your monitor, your lighting, your eyes. None of the three is the specified context of use.

Do

Record what you could not test, and treat it as a gap.

An untested assistive technology produced no findings, which is not the same as no defects — the distinction this whole system is built on.

Don't

Report "no accessibility issues found".

A checklist finds the defects it was written to look for. Saying so costs nothing; not saying so misleads whoever reads the file next.

NotJustAnyMed.Tech Design System · Accessibility checklist · v1.0 · draft for review
This is a design checklist, not a conformance audit and not legal advice.